| Do
I have to have a target density (Proctor, Marshall, Voidless) programmed
into the gauge in order for it to work?
No. The target value is only used by the gauge to calculate
the percent of compaction that has been achieved. The gauge will still
give Wet Density, Dry Density, Moisture and Percent Moisture results
without the target density entered. The percent compaction can be calculated
at a later time if necessary.
Example:
(DD/Proctor) x 100 = %Proctor or (WD/Marshall)
x 100 = %Marshall
› Top |
| |
| How
will frost (frozen material) affect the gauge readings?
The percent moisture result will be elevated and the dry density read
by the nuclear density gauge will be lower if the material contains
frost. Therefore, the tests will fail. When the water freezes it expands
and takes up a larger volume in the material beneath the gauge. Because
water is less dense than the measured material, the density will be
lower and the percent moisture will be higher. It is commonly written
in engineering specifications that "no fill may be placed over
frozen material". In extreme cases, settling could occur if frozen
material is covered and thaws at a later time.
› Top
|
| |
| My
gauge readings do not match my sand cone readings.
Usually they will not. There is a lot of room for operator
error when performing a sand cone test. There is really no way to prove
which test is more "accurate" because there is no absolute
test for soil density. The best way to compare the two test methods
is to take at least three 4 minute nuclear gauge measurements at different
locations (preferably direct transmission @ 6 or 8 inches) and at least
3 sand cone tests (at the same locations as the gauge readings). Find
the difference at all locations and average them.
› Top
|
| |
| Should
I take a backscatter or direct transmission reading on aggregate base?
(usually 4"- 6" in thickness)
Either one would be acceptable. The best thing to do is try
one of each. Ideally the direct transmission will give a more representative
reading, but it may also disturb the larger aggregate and lower the
density. If you try both in the same spot (backscatter test first) and
see which gives a better reading you can make the decision for yourself.
If the material thickness is over 4", direct transmission is more
appropriate in order to get a representative measurement.
› Top
|
| |
| What
is the prescale factor for my Troxler gauge?
In order to scale the count values to a reasonable number of
digits, Troxler devices use prescale factors when performing measurements,
standard counts and any other diagnostics count. In other words,
the count that a gauge displays is the total number of ionization events
measured by the gauge divided by the appropriate prescale factor.
The prescale factor is typically not needed by the gauge operator.
However, the ASTM standards D 6938 and D 2950 give a formula, to check
the stability of the standard counts, which requires the use of the
prescale factor. The prescale factors are listed below:
For Troxler Models 3401, 3411,
3430, 3430
Plus, 3440,
3440 Plus,
and 4300
devices, a prescale of 16 is to be used in the formula when a 4-minute
standard count is performed.
For Troxler Models 3450
and 4640
devices, a prescale of 8 is to be used in the formula when a 4-minute
standard count is performed.
› Top |
| |
| Can
the gauge read the moisture of a concrete slab?
It can read the moisture as a relative reading; in other words
it will give an indication of a problem area, but will not give a definite
moisture content. There are constituents in concrete that can cause
the moisture readings to be in error, but as long as the concrete is
homogeneous, the error will be constant. Example: If the slab reads
approximately 12% moisture in all areas but one area is approx. 20%,
it can be assumed that this is a problem area. This will give an indication
of where a core sample should be taken for investigative purposes.
› Top
|
| |
| My
3411 gauge is giving the ERROR 30 message.
This message means that the gauge cannot calculate
the test results. Usually this is due to the fact that there is no standard
count in the memory. Also, this could mean that the depth switch is
not correctly set; if the depth switch is set at a value much different
that that which is actually being measured, the calculated results will
be too large for the gauge to calculate. Try taking a new standard and
checking the depth switch before taking a measurement.
› Top
|
| |
| Why
is my gauge moisture result higher/lower than the oven dry/speedy moisture
results?
The oven dry tells the true moisture content by removing all
water from the sample. The gauge measures the hydrogen present in the
material; which is usually in the form of water. If the material contains
naturally occurring hydrogen or bound hydrogen, the gauge will measure
the moisture falsely high in many cases. (Some of these materials are:
mica, lime, fly ash, cement, organic materials, gypsum, coal, phosphates,
etc.) A false low reading can also occur but is less common (causes
include: high salt or iron oxide content or boron, lithium or cadmium
present). A moisture offset will adjust for this problem.
› Top
|
| |
| How
often do moisture/density gauges need to be calibrated?
We recommend that the gauges be properly evaluated for accuracy
or calibrated every 12 months.
› Top
|
| |
| What
are the ASTM and AASHTO standards written specifically for moisture
and density testing?
| • |
ASTM D 6938 & AASHTO T 310 In-Place Density
and Moisture Content of Soil and Soil-Aggregate by Nuclear Methods
(Shallow Depth) |
| • |
ASTM C 1040 Density of Unhardened and
Hardened Concrete In Place by Nuclear Methods |
| • |
ASTM D 2950 Density of Bituminous Concrete In Place by Nuclear
Methods |
› Top |
| |
| My
gauge is not charging. What could it be?
Reset
battery life to 190 hrs. Ensure charger is good. Check charger
jack to see if it is loose/broken. Possible bad battery pack.
› Top |
| |
| My
gauge displays tube error. What's the problem?
Take the gauge to a clean, dry environment and remove the control
panel or "scaler". Check for loose hardware inside the gauge
that could short the high voltage. Check for moisture inside the gauge
and, if moisture is present, remove the scaler and dry overnight with
a fan.
› Top
|
| |
| Why
does my gauge fail the standard count?
Ensure that the gauge is positioned
properly. Make sure that those standard counts stored are
appropriate by comparing to the factory standard counts reported in
the most recent calibration printout. If a standard count has not been
performed in more than a month and a new count fails by a small percentage
(less than 5%), accept the new count and perform four more. The fifth
standard count should pass.
› Top
|
| |
| When
I try to get my density, I get nothing but +'s. Why?
Obtain the gauge's most recent calibration printout and check
that the calibration constants in the memory match those on the printout.
Also, verify that the depth strip is functioning correctly. Check for
correct standard counts.
› Top
|
| |
| Why
are my gauge readings off?
Ensure you have a good standard count. Verify you are at the
correct depth. Check to make sure no offsets are enabled. Make sure
that the handle is clicked into the notch (depth position) properly.
Do not pass notch or push up on the handle after positioning it.
› Top
|
| |
| Why
is my gauge hard to index?
Magnalube is the recommended lubricant. Place a small amount
on the source rod and exercise the rod up and down. Check the bottom
of the gauge to make sure dirt has not accumulated in the sliding block
or base plate area.
› Top
|
| |
| Why
does my gauge indicate safe position when it is in BS?
You need to calibrate
the depth strip. If this doesn't fix the problem, your gauge
may need servicing.
› Top
|
| |
| My
gauge is locked up and there is no keypad response.
Press and hold the OFF button for 15-30 seconds until the gauge
turns off. If it does not turn off, remove the scaler and unplug the
ribbon cable from the baseboard. Plug the cable back in and turn the
gauge on allowing it to go through its 5 minute self-test.
› Top
|
| |
| My
3440/4640B is displaying random characters on the display.
Perform memory reset on gauge. 3440
reset / 4640
reset
› Top
|
| |
| Do
I need a MSDS for my nuclear gauges?
No. Troxler provides a MSDS
disclaimer statement for your files.
› Top
|
| |
| Which
USB devices are compatible with my Model 3440 Plus gauge?
A list of compatible USB devices is available here
. |
| |
| »
Return to top |
| |
| › Field
Equipment Main Page |
| › FAQs
Main Page |