| Precision
@ 6% asphalt |
| Sample |
1 min. |
4 min. |
8 min. |
16 min. |
| 7000 gram |
±0.084% |
±0.042% |
±0.029% |
±0.021% |
| 4" compacted |
±0.36% |
±0.18% |
±0.13% |
±0.09% |
| 6 " compacted |
±0.28% |
±0.14% |
±0.10% |
±0.07% |
|
 |
| Radiological |
Neutron Source |
100 µCi ±10% Cf-252 |
Source Form |
Stainless Steel Encapsulation, Special form |
Shielding |
Polyethylene and Cadmium |
Shipping Case |
DOT 7A, Type A |
 |
| Electrical |
Power Source |
110/220 V ac, 50/60 Hz; 12V vehicle battery or Alkaline
Batteries (D Cell) |
| Power Consumption |
1 Watt (nominal) |
| Baud Rate Range |
300 - 2400 baud |
| Test Data Storage |
Up to 99 test |
| Calibration Storage |
Up to 64 calibrations |
| Interface |
RS-232C for transfer to printer or computer |
 |
| Mechanical |
Length |
14.25 in. (36.2 cm) |
| Width |
11.00 in. (27.9 cm) |
| Height |
10.50 in. (26.7 cm) |
| Weight |
30.0 lbs. (13.62 kg) |
| Operational Temp. |
0 to 140ºF (-18 to 60ºC) |
| Sample Temp. |
0 to 350ºF (-18 to 177ºC) |
 |
| Mechanical
(Control Unit) |
Length |
8.62 in. (21.9 cm) |
| Width |
11.00 in. (27.9 cm) |
| Height |
3.60 in. (9.2 cm) |
| Weight |
2.75 lbs. (1.25 kg) |
| Operational Temp. |
0 to 140ºF (-18 to 60ºC) |
| Sample Temp. |
0 to 350ºF (-18 to 177ºC) |
 |
| • |
Four stainless steel pans supplied. |
| • |
One 4" fixture supplied with Universal Sample System.
(6" available as an option.) |
| • |
Four Helium3 thermal neutron detectors supplied. |
| • |
Hardshell case for transport supplied. |
|